1.1 This practice covers the description of requirements of spectrophotometric performance, especially for test methods, and the testing of the adequacy of available equipment for a specific method (for example, qualification for a given application). The tests give a measurement of some of the important parameters controlling results obtained in spectrophotometric methods, but it is specifically not to be concluded that all the factors in instrument performance are measured, or in fact may be required for a given application.
1.1.1 This practice is primarily directed to dispersive spectrophotometers used for transmittance measurements rather than instruments designed for diffuse transmission and diffuse reflection.
| Author | ASTM |
|---|---|
| Editor | ASTM |
| Document type | Standard |
| Format | File |
| Confirmation date | 2013-01-01 |
| ICS | 17.180.30 : Optical measuring instruments |
| Number of pages | 11 |
| Replace | ASTM E275-08 |
| Set | ASTMVOL0306 |
| Year | 2008 |
| Document history | ASTM E275-08 |
| Country | USA |
| Keyword | ASTM 275;ASTM E275;ASTM E275;10.1520/E0275-08R13 |