Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems

More details

$26.66

-57%

$62.00

More info

1.1 This practice addresses the techniques of measurement and reporting of basic static performance (accuracy, repeatability, and so forth) of surgical navigation and/or robotic positioning devices under defined conditions. The scope covers the tracking subsystem, testing only in this practice the accuracy and repeatability of the system to locate individual points in space. A point in space has no orientation; only multi-dimensional objects have orientation. Therefore, orientation of objects is not within the scope of this practice. However, in localizing a point the different orientations of the localization tool can produce errors. These errors and the orientation of the localization tool are within the scope of this practice. The aim is to provide a standardized measurement of performance variables by which end-users can compare within (for example, different fixed reference frames or stylus tools) and between (for example, different manufacturers) different systems. Parameters to be evaluated include (based upon the features of the system being evaluated):

(1) Location of a point relative to a coordinate system.

(2) Relative point to point accuracy (linear).

(3) Repeatability of coordinates of a single point.

(4) For an optically based system, the range of visible orientations of the reference frames or tools.

(5) This method covers all configurations of tool arrays in the system.

Author ASTM
Editor ASTM
Document type Standard
Format File
ICS 35.240.80 : IT applications in health care technology
Number of pages 11
Replace ASTM F2554-10 + Redline
Set ASTMVOL1302
Year 2018
Document history ASTM F2554-10 + Redline
Country USA
Keyword ASTM 2554;ASTM F2554;ASTM F2554;10.1520/F2554-18